Diagnostic function aiming at metrological parameters determination (repeatability), the parameters are determined for the actual conditions of use. When speaking of repeatability it may be also used for weighing time optimization. Autotest is operated in an automatic mode thus operator’s time is saved.
Radwag XA 5Y.M.A Microbalance
Regular price
$14,161.00
Sale price
$14,161.00
Regular price
Unit price
About this Model:
Shipping Lead Time: 4 - 6 Weeks

Radwag XA 5Y.M.A Microbalance
Radwag XA 5Y.M.A Microbalances






Radwag XA 5Y.M.A Microbalances






Datasheet | |
---|---|
Maximum capacity [Max] | 6.1 g |
Minimum load | 0.1 mg |
Readability [d] | 1 µg |
Verification unit [e] | 1 mg |
Tare range | -6.1 g |
Standard repeatability [5% Max] | 0.8 µg |
Standard repeatability [Max] | 2.5 µg |
Standard minimum weight (USP) | 1.6 mg |
Standard minimum weight (U=1%, k=2) | 0.16 mg |
Permissible repeatability [5% Max] | 1.5 µg |
Permissible repeatability [Max] | 3 µg |
Linearity | ±7 µg |
Eccentric load deviation | 7 µg |
Sensitivity time drift | 1×10⁻⁶/Year×Rt |
Stabilization time | ~3.5 s |
Adjustment | internal (automatic) |
OIML Class | I |
Leveling system | automatic - Reflex Level System |
Display | 10″ touchscreen |
Weighing chamber doors | automatic |
Delivery components | Microbalance, weighing pan, weighing pan shield, power supply, brush, fabric dust cover. |
Weighing chamber dimensions | 199×170×217 mm |
Weighing pan dimensions | ø30 mm |
Packaging dimensions | 750×492×595 mm |
Net weight | 14.5 kg |
Gross weight | 18.9 kg |
Protection class | IP 43 |
Relative humidity | 40% — 80% |
Communication interface | USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot |
Power supply | Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A Balance: 12 – 15V DC 1.4A max* |
Operating temperature | +10 — +40 °C |
Operating temperature change rate | ±0.3°C/1h (±1°C/8h) |
Relative humidity change rate | ±1%/h (±4%/8h) |